
Appendix D. CNF4 Heater/Ventilator
D.3 Wiring
The following table shows the recommended datalogger wiring for using the
CNR4 sensor with the CNF4 heater/ventilator while making the differential
measurement.
TABLE D-1. CR1000 and CR3000 Datalogger Connections for
Differential Measurement with Heater/Ventilator Control
Function Wire Color CR6, CR1000, CR3000
Pyranometer Up Signal Red Differential Input (H)
Pyranometer Down Signal White Differential Input (H)
Pyranometer Down Reference *Black Differential Input (L)
Pyrgeometer Up Signal Grey Differential Input (H)
Pyrgeometer Down Signal Brown Differential Input (H)
Pyrgeometer Down Reference *Green Differential Input (L)
Shield Clear
Thermistor Signal White Single-Ended Input
Thermistor Voltage Excitation Red
Voltage Excitation
(VX)
Thermistor Signal Reference Black
Shield Clear
CNF4 Heater/Ventilator
Ventilator Ground Blue G
Heater Power Green SW12V
Heater Ground Yellow G
*Jumper to with user supplied wire
Pull back wires for Pt-100 (grey, brown, green, and yellow), which are not in
use, and tie them around the TEMP cable using a cable tie or electrical tape to
avoid possible damage to the Pt-100, due to electrical short circuit.
Do not use the datalogger’s switched 12 V to simultaneously
power the heater and ventilator. Simultaneously powering
the heater and ventilator will exceed the current limit of the
switched 12 V channel. If the heater and ventilator need to
be used at the same time, connect the CNF4 to the 12V
channel instead of the switched 12 V channel and use an
external relay to switch the power on and off. Refer to
Section 4.1 of the CR1000 and CR3000 manual for details
on the 12V current source limits.
D-7
Komentáře k této Příručce