Campbell CR800 Specifikace Strana 141

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 324
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 140
Section 6. Data Table Declarations and Output Processing Instructions
6-23
The second dimension, when greater than 1, is determined by the value of the
element in the source array following the element checked for the crossing. It
is the value of this variable at the time the crossings are detected that
determines the second dimension.
0
0.5
1
1.5
2
2.5
3
3.5
0123456
Crossing Levels
Crossing Source
2nd Dim Boundary
2ndDim Source
FIGURE 6.4-2. Crossing Data with Second Dimension Value
Point Crossing
Source
2nd Dim
Source
Action
1 0.5 .7 First value, no counts
2 1.2 1.8 Add one count to first crossing,
second 2D bin, the signal crossed
1
Histogram:
2D < 1.25 1.25<2D<2.25 2.25<2D<3.25
Cross 1
0 1 0
Cross 1.5
0 0 0
Cross 3
0 0 0
3 1.4 .7 No levels crossed, no counts
4 0.3 .7 Crossed a level but was falling
edge, no counts
5 3.3 2.7 Add one count to first, second, and
third crossing bins in the third 2D
bin, the signal crossed 1, 1.5 and
3.