Campbell CS106 Manuál s instrukcemi Strana 17

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 28
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 16
CS106 Barometric Pressure Sensor
TABLE 7-1. Signal and Ground Connectors for CS106
Wire
CS106 Terminal
Datalogger
Single-Ended Measurement
Datalogger
Differential Measurement
Blue VOUT S.E. Input High Side of Diff Input
Yellow AGND
Low Side of Diff. Input
Black GND CR9000(X)
G (Other Dataloggers)
(CR9000(X)
G (Other Dataloggers)
Green EXT TRIG Control port (use to turn power
on/off)
Control port (use to turn
power on/off)
Red SUPPLY 12 Vdc 12 Vdc
Shield Shield
Improper wiring may damage the CS106 beyond repair.
7.3.2 5-pin Screw Terminal Plug Connector
The datalogger connects to the CS106 via a 5-pin screw terminal plug
connector. This connector is removable and may be replaced. The
replacement connector may come with a connector key attached to it to ensure
that the connector is plugged into the CS106 right side up (see FIGURE 7-4).
When the connector is right side up, it will easily plug into the barometer.
FIGURE 7-4. Connector key attached to 5-pin screw terminal plug
connector
A 5-pin screw terminal that is plugged in upside down
will damage the sensorperhaps beyond repair.
WARNING
WARNING
9
Zobrazit stránku 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 27 28

Komentáře k této Příručce

Žádné komentáře