OBS-3A Turbidity and
Temperature Monitoring System
Revision: 11/11
Copyright © 2007-2011
Campbell Scientific, Inc.
Zobrazit stránku 0
1 2 3 4 5 6 ... 57 58

Shrnutí obsahu

Strany 1 - Temperature Monitoring System

OBS-3A Turbidity and Temperature Monitoring SystemRevision: 11/11Copyright © 2007-2011Campbell Scientific, Inc.

Strany 2 - WARRANTY AND ASSISTANCE

Section 2. Instrument Setup Surveys The OBS-3A will usually be towed from a cable harness for surveys. The serial cable supplied with the unit is st

Strany 3 - PLEASE READ FIRST

Section 2. Instrument Setup For extended deployment time, lithium batteries are a good alternative to alkaline batteries. Campbell Scientific sells

Strany 4

Section 2. Instrument Setup 2-4

Strany 5 - OBS-3A Table of Contents

Section 3. Operations 3.1 Software Installation Insert the CD and select “Install OBS-3A Utility”. Follow the installation wizard to install the so

Strany 6

Section 3. Operations Data received from OBS-3A while it is connected to the PC will be stored in this file (see FIGURE 3.2-3). FIGURE 3.2-3. Dat

Strany 7 - Section 1. Introduction

Section 3. Operations 4. Click Connect/Disconnect to get a green light and synchronize the OBS-3A clock with your PC by clicking . 3.3 Pull-Down M

Strany 8 - 1.1 OBS Sensor

Section 3. Operations box to select ONLY change host computer port. Then click Apply and the button. If you get the OBS-3A information box, the bau

Strany 9 - Section 2. Instrument Setup

Section 3. Operations FIGURE 3.5-1. Test Data Sample FIGURE 3.5-2. Window for Viewing Instrument Information 3-5

Strany 10 - 2.2 Battery Installation

Section 3. Operations 3.6 Water-Density and Barometric Corrections Since depths are estimated from pressure measurements, it is important to set the

Strany 11

Section 3. Operations length or the 2048 / rate. Note: the product of the rate and the duration cannot exceed 2048. Rate: Rate is the frequency of

Strany 12

WARRANTY AND ASSISTANCE This equipment is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. (“CSC”) to be free from defects in materials and wor

Strany 13 - Section 3. Operations

Section 3. Operations interval that will show the changes in turbidity and water depth that you wish to investigate. Rate sets the number measuremen

Strany 14

Section 3. Operations 3.11 Surveying If you have a pressure sensor, click the OBS menu and select Barometric Correction (do not do this when the OBS-

Strany 15 - 3.4 Communication Settings

Section 3. Operations 4. A file for logging data was created when you started the OBS-3A Utility. You can review data at any time with Open and imp

Strany 16 - 3.5 Testing Sensors

Section 3. Operations 3.13 Scheduled Sampling Use this mode when you want the OBS-3A to sample at specific times, scheduled by hours and minutes, on

Strany 17

Section 3. Operations 4. Select setpoint values for transitions to fast sampling (SLOW>>>FAST) and slow (FAST>>>SLOW) rates. 5.

Strany 18 - 3.8 Definitions

Section 3. Operations 7. Wait for the progress bar to disappear and examine data as a plot or test file (Section 3.17—Graphing and Printing). 3.16 S

Strany 19 - 3.9 Sampling Schedules

Section 3. Operations 3.18 Excel Spreadsheets To make an Excel spreadsheet from OBS-3A data, start Excel and set file type to All. Open a data file a

Strany 20 - 3.10 Sampling Modes

Section 3. Operations 3.19 Erasing Data Memory To erase the flash data memory, do the following: 1) Click on the Terminal Mode Icon. 2) At the OB

Strany 21 - 3.11 Surveying

Section 3. Operations 3-16

Strany 22 - 3.12 Cyclic Sampling

Section 4. Troubleshooting This section will help you isolate problems that can be easily fixed such as cable-continuity, processor reset, and batte

Strany 23 - 3.14 Setpoint Sampling

PLEASE READ FIRST About this manual Please note that this manual was originally produced by Campbell Scientific Inc. (CSI) primarily for the US market

Strany 24 - 3.15 Data Retrieval

Section 4. Troubleshooting 4. The OBS-3A and PC are not set to the same baud rate or communication protocol (e.g. RS-232, USB, RS-485). x Click an

Strany 25 - 3.17 Graphing and Printing

Section 4. Troubleshooting Bright sun near the surface ( < 2 meters) or black-colored sediments cause erroneous OBS readings. Do not survey in sh

Strany 26 - 3.18 Excel Spreadsheets

Section 4. Troubleshooting 4-4

Strany 27 - 3.19 Erasing Data Memory

Section 5. Maintenance 5.1 OBS Sensor The OBS sensor must be kept clean to measure sediment concentration or turbidity accurately. A gradual declin

Strany 28

Section 5. Maintenance during surveys or after each deployment. A sensor that has been stored dry should be soaked in water for 15 minutes prior to

Strany 29 - Section 4. Troubleshooting

Section 5. Maintenance carefully inspected every six months and serviced before all deployments longer than one month. 1. Disassemble O-ring seals a

Strany 30

Section 5. Maintenance 5-4

Strany 31

Section 6. Calibration 6.1 Turbidity This section briefly describes the materials and equipment you will need and the basic procedures for calibrati

Strany 32

Section 6. Calibration 6.1.2 Preparation 1. Experience has shown that only three calibration points are needed to get sub-1% accuracy. It is recomm

Strany 33 - Section 5. Maintenance

Section 6. Calibration There is no way to cancel a calibration after the Compute button is clicked. WARNING Review the quality assurance checks discu

Strany 35 - 5.7 User-serviceable Parts

Section 6. Calibration Where: Tstd = Turbidity of the standard solution; Tstk = Turbidity of the stock solution, usually 4000 NTU; Vstk = Cumulative

Strany 36

Section 6. Calibration x 2 gallons filtered distilled water (purified water from the super market works fine), x Hand-drill motor, x Paint stirrer

Strany 37 - Section 6. Calibration

Section 6. Calibration Preparation 1. Clean containers and glassware with detergent and rinse with filtered water. 2. Perform the calibration under

Strany 38 - 6.1.2 Preparation

Section 6. Calibration 9. Weigh 5 to l0 equal increments of the sediment so that the total dry weight will produce the maximum concentration expecte

Strany 39

Section 6. Calibration 6-8

Strany 40 - 6.2 Sediment

Section 7. Optics and Turbidity Measurements Turbidity is the cloudy appearance of a liquid produced by light scattered from suspended matter. It i

Strany 41 - 6.2.2 Sediment Preparation

Section 7. Optics and Turbidity Measurements 7-2 Can turbidity be converted to suspended solids concentrations and vise-versa? In most situations, co

Strany 42 - )(1 i kgVu

Section 8. Factors Affecting OBS Response This section summarizes some of the important factors that affect OBS-3A measurements and shows how ignori

Strany 43 - Calibrations

Section 8. Factors Affecting OBS Response 8.2 Suspensions with Mud and Sand As mentioned earlier, backscattering from particles is inversely related

Strany 44

Section 8. Factors Affecting OBS Response 0510Sediment Concentration (g/l)15012345Signal (V)A FIGURE 8.3-1. Response at High Sediment Concentrations

Strany 45 - Measurements

OBS-3A Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links

Strany 46

Section 8. Factors Affecting OBS Response 02468Munsell Value (Black = 0)100.00.20.40.60.81.0Infrared ReflectanceCalciteBytowniteActinoliteMagnetite F

Strany 47 - Response

Section 8. Factors Affecting OBS Response in this region. OBS sensors are also more sensitive to mineral particles than either bubbles or particulat

Strany 48 - Gain (V/g/l)

Section 8. Factors Affecting OBS Response 8-6

Strany 49 - 8.4 Sediment Color

Section 9. References See www.campbellsci.com/obs for a complete list of references. Conner, C.S. and A.M. De Visser. 1992. A Laboratory Investigati

Strany 50 - 8.6 Bubbles

Section 9. References 9-2 U.S. Geological Survey. 2003. National Field Manual of the Collection of Water-Quality Data. Book 9, Handbooks for Water-Re

Strany 51 - Backscatter (90 - 180

Section 10. Specifications MEASUREMENT RANGE Turbidity (AMCO Clear)... 0.4 to 4,000 NTU1 Mud (D50=20μm) ...

Strany 52

Section 10. Specifications 10-2 CSI # 21135 End Cap Screws, 8-32 x 3/8”, socket (5/64”) CSI # 21120 Dummy Plug CSI # 21122 Plug Locking Sleeve, Subco

Strany 54

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Strany 55 - Section 10. Specifications

OBS-3A Table of Contents ii 6.1.3 OBS-3A Utility Software Steps... 6-2 6.1.4 Making Turbidity Standards

Strany 56

Section 1. Introduction The heart of the OBS-3A monitor is an OBS sensor for measuring turbidity and suspended solids concentrations by detecting n

Strany 57

Section 1. Introduction Depending on the number of sensors and the statistics selected, the OBS-3A can log as many as 200,000 lines of data (one per

Strany 58

Section 2. Instrument Setup 2.1 Mounting Suggestions Maximum depth for the OBS-3A housing is 300 meters. Working depths for individual instruments