OBS-3A Turbidity and Temperature Monitoring SystemRevision: 7/13Copyright © 2007-2013Campbell Scientific, Inc.

Strany 2

OBS-3A Turbidity and Temperature Monitoring System 3.1 Ships With CSI pn 21229 Accessory Kit ResourceDVD CSI pn 29225 HydroSci Software on DVD 4. Ov

Strany 3 - Warranty

OBS-3A Turbidity and Temperature Monitoring System FIGURE 4-1. Anatomy of an OBS sensor 4.2 Temperature and Optional Sensors Temperature is measure

Strany 4 - Assistance

OBS-3A Turbidity and Temperature Monitoring System Light transmission in water is attenuated by scattering (deflection by water molecules, and suspend

Strany 5 - Table of Contents

OBS-3A Turbidity and Temperature Monitoring System 5.1 Measurement Range Turbidity (AMCO Clear): 0.4 to 4,000 NTU1 Mud (D50=20 µm): 0.4 to 5,000 mg/

Strany 6

OBS-3A Turbidity and Temperature Monitoring System Battery capacity: 18 A h Maximum battery life: 8,000 hours Data protocols: RS-232 / RS-485 Maximu

Strany 7

OBS-3A Turbidity and Temperature Monitoring System TABLE 6-1. Working and Maximum Depths Pressure Sensor Working Depth Maximum Depth 0.2 Bar 0 to

Strany 8

OBS-3A Turbidity and Temperature Monitoring System the cable grip and connector and put chafe protection on the sensor head where it contacts the wire

Strany 9 - Monitoring System

OBS-3A Turbidity and Temperature Monitoring System FIGURE 6-2. Battery installation For extended deployment time, lithium batteries are a good alter

Strany 10 - 4. Overview

OBS-3A Turbidity and Temperature Monitoring System 6.3 Running HydroSci 1. Select the HydroSci program to start the program. 2. Physically connect

Strany 11

OBS-3A Turbidity and Temperature Monitoring System 3. Select OBS-3A on the lift side of the screen and select the appropriate COM Port and Baud Rate

Strany 13 - 5.4 Other Data

OBS-3A Turbidity and Temperature Monitoring System 7. After you have finished interacting with your sensor, click on the Connection tab and press the

Strany 14 - 6. Operations

OBS-3A Turbidity and Temperature Monitoring System Blow on the temperature sensor to observe an increase in temperature (red trace line on the top plo

Strany 15

OBS-3A Turbidity and Temperature Monitoring System Show Terminal tab brings up a terminal screen that allows you to view the commands being sent to th

Strany 16 - 6.1.2 Battery Installation

OBS-3A Turbidity and Temperature Monitoring System 6.8.2 Operations Put OBS-3A to Sleep Press the Sleep button to put the OBS-3A in low-power, slee

Strany 17 - 6.2 Software Installation

OBS-3A Turbidity and Temperature Monitoring System HydroSci will confirm if you want to put the OBS-3A to sleep. Clear OBS-3A Data Storage Press the

Strany 18 - 6.3 Running HydroSci

OBS-3A Turbidity and Temperature Monitoring System Barometric Correction The OBS-3A measures absolute pressure so a correction must be made for barome

Strany 19

OBS-3A Turbidity and Temperature Monitoring System Change to RS-232 To switch from RS-485 communication to RS-232 communication, press the RS-232 butt

Strany 20 - 6.4 Testing Sensors

OBS-3A Turbidity and Temperature Monitoring System If it becomes necessary to load coefficients from a saved file, press the Load Coefficients from Fi

Strany 21 - 6.6 Menus

OBS-3A Turbidity and Temperature Monitoring System Calibration The Calibration box can be used to calibrate the OBS-3A for NTU or mg/L measurements.

Strany 22 - 6.8 OBS-3A Configuration

OBS-3A Turbidity and Temperature Monitoring System You will be asked whether or not you want to log data to a file on your computer. In survey mod

Strany 23 - 6.8.2 Operations

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Strany 24

OBS-3A Turbidity and Temperature Monitoring System You can press the Load Configuration From File button to load a configuration that has previously

Strany 25

OBS-3A Turbidity and Temperature Monitoring System Units of Measure Select the units for depth (Meters or Feet). Sensor Measurements Select the check

Strany 26

OBS-3A Turbidity and Temperature Monitoring System You will be asked whether or not you want to log data to a file on your computer. After setting

Strany 27

OBS-3A Turbidity and Temperature Monitoring System • Wave Setup Wave Setup fields are only enabled when Depth is selected under Sensor Measurements a

Strany 28 - 6.8.3 Survey Configuration

OBS-3A Turbidity and Temperature Monitoring System o Rate Rate is the frequency of sampling for the duration of measurements. All sensors are samp

Strany 29

OBS-3A Turbidity and Temperature Monitoring System Go to the Connection tab and press Disconnect to disconnect from the OBS-3A. Unplug the test cabl

Strany 30

OBS-3A Turbidity and Temperature Monitoring System • Water Density Specify the water temperature and salinity. The specified temperature and salini

Strany 31 - 6.8.4 Cyclic Configuration

OBS-3A Turbidity and Temperature Monitoring System • Sample Settings o Duration This is the length of time in seconds that the OBS-3A is measuring

Strany 32

OBS-3A Turbidity and Temperature Monitoring System 6.8.6 Setpoint Configuration Setpoint Configuration mode is used for fast sampling of events such

Strany 33

OBS-3A Turbidity and Temperature Monitoring System • Sensor Measurements Select the check boxes next to the sensor measurements you wish to make. F

Strany 34

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 35

OBS-3A Turbidity and Temperature Monitoring System • Interval Settings o Slow Interval The time, in seconds, between the start of one sample and t

Strany 36

OBS-3A Turbidity and Temperature Monitoring System Temperature, Salinity, Concentration, and Depth are shown on the charts. Only the sensor measureme

Strany 37

OBS-3A Turbidity and Temperature Monitoring System All sensor measurements selected in the configuration will be shown in the table view. Use the scr

Strany 38

OBS-3A Turbidity and Temperature Monitoring System 6.10 View Data The View Data screen can be used to view data stored on the OBS-3A from cyclic, sch

Strany 39

OBS-3A Turbidity and Temperature Monitoring System 3. A screen asking if the OBS-3A is currently sampling will appear. If you are ready to stop samp

Strany 40 - 6.9 Monitor

OBS-3A Turbidity and Temperature Monitoring System 4. Go to View Data to save data in a file. 5. Highlight the data with the start and end times y

Strany 41

OBS-3A Turbidity and Temperature Monitoring System 6. Once the correct data is selected, press Save Data File. The Save As screen will appear, name

Strany 42

OBS-3A Turbidity and Temperature Monitoring System Then press the New Line Graph button and a new screen will appear with various graphing options.

Strany 43 - 6.10 View Data

OBS-3A Turbidity and Temperature Monitoring System . The spreadsheet of data can also be printed. 6.11 Show Terminal The Show Terminal tab brings u

Strany 44

OBS-3A Turbidity and Temperature Monitoring System 6.12 Shutdown From the OBS-3A Configuration | Operations tab, select Put OBS-3A To Sleep. It wi

Strany 45

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Strany 46

OBS-3A Turbidity and Temperature Monitoring System 6.14 Erasing Memory Data To erase the memory data on the OBS-3A, go to OBS-3A Configuration | Op

Strany 47

OBS-3A Turbidity and Temperature Monitoring System 7. Calibration It is a good idea to use the Retrieve and Save Coefficients to File button (under

Strany 48 - 6.11 Show Terminal

OBS-3A Turbidity and Temperature Monitoring System • 100 ml to deliver (TD) volumetric pipette • 25 ml TD measuring pipette • 2 gallons filtered di

Strany 49 - 6.13 Excel Spreadsheets

OBS-3A Turbidity and Temperature Monitoring System 3. Prepare the next NTU standard and put the OBS-3A in it. Repeat steps 1 and 2 for all calibrati

Strany 50 - 6.14 Erasing Memory Data

OBS-3A Turbidity and Temperature Monitoring System The formula for preparing turbidity standards other than shown in TABLE 7-1 is: ⎥⎦⎤⎢⎣⎡+=stkdwstkstk

Strany 51 - 7. Calibration

OBS-3A Turbidity and Temperature Monitoring System 7.2.1 Equipment and Materials • Dry, completely disaggregated bottom sediment or suspended matter

Strany 52 - 7.1.2 Preparation

OBS-3A Turbidity and Temperature Monitoring System 10 20 30 40 50Sediment Concentration (mg/l)0102030405060Turbidity (NTU)Hand Shaking15-min. Sonic Ba

Strany 53

OBS-3A Turbidity and Temperature Monitoring System Sediment concentrations are calculated with the following equations: mg/lρMVMssis=⎥⎦⎤⎢⎣⎡+ ;

Strany 54 - 7.2 Sediment

OBS-3A Turbidity and Temperature Monitoring System FIGURE 8-1. Component locations Unit does not communicate with PC. There are several possible c

Strany 55 - 7.2.2 Sediment Preparation

OBS-3A Turbidity and Temperature Monitoring System b. If the OBS-3A still fails to respond, try changing baud rate speeds by selecting Attempt to f

Strany 56 - 3-min. Sonic Probe

Table of Contents 6.13 Excel Spreadsheets ... 41 6.14 Erasing Memory Data..

Strany 57 - 8. Troubleshooting

OBS-3A Turbidity and Temperature Monitoring System Power failed due to battery clip corrosion or a broken power wire. Check for a broken red wire conn

Strany 58

OBS-3A Turbidity and Temperature Monitoring System 9. Maintenance 9.1 OBS Sensor The OBS sensor must be kept clean to measure sediment concentration

Strany 59

OBS-3A Turbidity and Temperature Monitoring System during surveys or after each deployment. A sensor that has been stored dry should be soaked in wat

Strany 60

OBS-3A Turbidity and Temperature Monitoring System 9.5 Pressure Housing The pressure housing and O-ring seals require little maintenance unless the h

Strany 61 - 9. Maintenance

OBS-3A Turbidity and Temperature Monitoring System 10.1 Particle Size The size of suspended sediment particles typically ranges from about 0.2 to 500

Strany 62 - 9.4 Batteries

OBS-3A Turbidity and Temperature Monitoring System 10.2 Suspensions with Mud and Sand As mentioned earlier, backscattering from particles is inversel

Strany 63 - 9.7 User-serviceable Parts

OBS-3A Turbidity and Temperature Monitoring System 10.3 High Sediment Concentrations At high sediment concentrations, particularly in suspensions of

Strany 64 - 10.1 Particle Size

OBS-3A Turbidity and Temperature Monitoring System 10.4 Sediment Color Sediment color, after particle size, has a major effect on OBS sensitivity, an

Strany 65 - Gain = 8.3(D)

OBS-3A Turbidity and Temperature Monitoring System 10.6 Bubbles Although bubbles efficiently scatter IR, monitoring in most natural environments show

Strany 66 - Signal (V)

OBS-3A Turbidity and Temperature Monitoring System 10.7 Biological and Chemical Fouling Sensor cleaning is essential during extended deployments. In

Strany 67 - 10.5 Water Color

Table of Contents Tables 6-1. Working and Maximum Depths...7 7-1. Mixing Volumes for Formazi

Strany 68 - 10.6 Bubbles

OBS-3A Turbidity and Temperature Monitoring System 62 Standard Methods for the Examination of Water and Wastewater, 20th Edition. 1998. 2130 Turbidity

Strany 70

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Strany 71

Table of Contents iv

Strany 72

OBS-3A Turbidity and Temperature Monitoring System 1. Introduction The OBS-3A combines our OBS® probe with pressure, temperature, and conductivity s