HMP155ATemperature and Relative Humidity Probe
The HMP155A can be powered continuously or the power may be switched to
conserve battery life. The HMP155A consumes less than 3 milliamps current
at 12 Vdc. Approximately 2 seconds is required for the sensor to warm up
after power is switched on. At measurement rates slower than once per 5
seconds, the overall power consumption (datalogger and sensors) may be
reduced by switching power to the HMP155A. Most current Campbell
Scientific dataloggers have a built-in switched 12 Vdc that can be used to
control power. The CR9000 and many Edlog dataloggers do not have a built-in
switched and require the SW12 to switch power (see at
www.campbellsci.com/old-manuals or contact a Campbell Scientific
application engineer for more information).
HMP155 sensors purchased directly from Vaisala with serial
numbers < E4430001 require approximately 5 seconds warm up
time.
The -L option on the model HMP155A Temperature and Relative Humidity
probe (HMP155A-L) indicates that the cable length is user specified. TABLE
5-1 gives the recommended lead length. Add 2 ft to the cable length if the
enclosure is mounted to the leg base of a CM106B, CM110, CM115, or
CM120 tripod.
TABLE 5-1. Recommended Lead Lengths
2 m Height Atop a tripod or tower via a 2 ft crossarm
Mast/Leg CM202 CM106B CM110 CM115 CM120 UT10 UT20 UT30
9 ft 11 ft 14 ft 14 ft 19 ft 24 ft 14 ft 24 ft 37 ft
The probe’s cable can terminate in:
Pigtails that connect directly to a Campbell Scientific datalogger
(option PT).
Connector that attaches to a prewired enclosure (option PW). Refer
to www.campbellsci.com/prewired-enclosures for more information.
6. Specifications
Features:
Well-suited for long-term, unattended applications
Accurate and rugged
Compatible with Campbell Scientific CRBasic dataloggers:
CR200(X) series, CR800 series, CR1000, CR3000, CR5000, and
CR9000(X).
Operating temperature range
for humidity measurement: 80 to 60 °C (–112 to 140 °F)
Storage temperature range: 80° to 60 °C (–112 to 140 °F)
NOTE