QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES
TITLE
SERVICING AND CALIBRATION OF OPTICAL MONITORING
DATALOGGERS
TYPE
STANDARD OPERATING PROCEDURE
NUMBER
4250
DATE
MARCH 1994
AUTHORIZATIONS
TITLE NAME SIGNATURE
ORIGINATOR James H. Wagner
PROJECT MANAGER James H. Wagner
PROGRAM MANAGER David L. Dietrich
QA MANAGER Gloria S. Mercer
OTHER
REVISION HISTORY
REVISION
NO.
CHANGE
DESCRIPTION
DATE AUTHORIZATIONS
1.0 Add responsibilities and equipment. May 1996
Zobrazit stránku 0
1 2 3 4 5 6 ... 55 56

Shrnutí obsahu

Strany 1

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF OPTICAL MONITORINGDATALOGGERSTYPESTANDARD OPERATING PROCEDUREN

Strany 2 - Page i of i

Number 4250-2000Revision 1.0Date MAY 1996Page 1 of 71.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing an

Strany 3 - Page 1 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 2 of 7• Ensure that all datalogger servicing is documented and archived in accordance with theprocedures

Strany 4 - Page 2 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 3 of 7• Replacement components as required• Medium screwdrivers (flat-blade and Phillips-head)• Battery

Strany 5 - Page 3 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 4 of 7CAMPBELL 21X DATALOGGERSERVICING DOCUMENTATION FORMDate: ___________________________Datalogger S/N

Strany 6 - Page 4 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 5 of 7After a few seconds delay, the memory check results will bedisplayed. If all memory is installed

Strany 7 - Page 5 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 6 of 7Documentation Form. The correct reading for each port is shown,along with the manufacturers'

Strany 8

Number 4250-2000Revision 1.0Date MAY 1996Page 7 of 7RECHARGE If the battery voltage is less than 11.76 volts, connect theBATTERY datalogger to the bat

Strany 9 - Page i of i

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF THE HANDAR 540A/570A DCPTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Strany 10 - Page 1 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page i of iiTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Man

Strany 11 - Page 2 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page ii of iiLIST OF FIGURES (CONTINUED)Figure Page4-2 Front Panel Configurations - Handar 540A/570A DCP 54-

Strany 12 - Page 3 of 7

Number 4250 Revision 1.0 Date MAY 1996 Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONS

Strany 13 - SERVICING DOCUMENTATION FORM

Number 4250-2010Revision 0.1Date MAY 1996Page 1 of 371.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing a

Strany 14 - Page 5 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 2 of 37• Coordinate with Handar for return, servicing, and recalibration of 540A/570A DCPsthat fail to o

Strany 15 - Page 6 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 3 of 37• Electronic contacts cleaning fluid• Medium screwdrivers (flat-blade and Phillips-head)4.0 METHO

Strany 16 - Page 7 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 4 of 37POST-FIELD INSPECTION CHECKLISTHANDAR 540A/570A DCPDCP S/N: ______________________________Site:

Strany 17

Number 4250-2010Revision 0.1Date MAY 1996Page 5 of 37Figure 4-2. Front Panel Configurations - Handar 540A/570A DCP.

Strany 18 - Page i of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 6 of 374.1.2 Physical Inspection - ExternalINSPECT Inspect the outside of the DCP case thoroughly. Look

Strany 19 - Page ii of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 7 of 37Figure 4-3. DCP Component Diagram.

Strany 20 - Page 1 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 8 of 37SWITCH When the "TERM" setup screen (see Figure 4-4) is displayed onBAUD RATE the compu

Strany 21 - Page 2 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 9 of 37Figure 4-4. TERM Setup Screen.

Strany 22 - Page 3 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 10 of 37Figure 4-5. "TERM" DCP Programming Screen (Program in DCP Memory).

Strany 23 - HANDAR 540A/570A DCP

Number 4250 Revision 1.0 Date MAY 1996 Page 1 of 51.0 PURPOSE AND APPLICABILITYThis standard operating procedure (SOP) ou

Strany 24 - Page 5 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 11 of 37Figure 4-6. TERM DCP Programming Screen (No Program in DCP Memory).

Strany 25 - Page 6 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 12 of 37SWITCH Switch the DCP from "RUN" mode to "PROGRAM" mode byMODES entering ?.T

Strany 26 - Page 7 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 13 of 37COMPARE The DCP data channel 04 reading (Rotronics AT output signal)READINGS must be adjusted by

Strany 27 - Page 8 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 14 of 37ROUTINE SERVICING CHECKLISTHANDAR 540A/570A DCPDCP S/N: __________________________Date: ________

Strany 28 - Page 9 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 15 of 37VERIFY Verify that the GOES primary channel select switches are set toGOES “900” and the seconda

Strany 29 - Page 10 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 16 of 37CLEAN INTERIOR Clean the inside of the DCP with compressed air.OF DCPMODIFY 570A If this is a Ha

Strany 30 - Page 11 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 17 of 37Figure 4-8. Handar 570A ADC Board - Component Locations.

Strany 31 - Page 12 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 18 of 37Figure 4-9. Handar 540A Met Board - Component Locations.

Strany 32 - Page 13 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 19 of 374.3 DCP PROGRAMMINGThe Handar 540A/570A DCP operational configuration is established through a u

Strany 33 - ROUTINE SERVICING CHECKLIST

Number 4250-2010Revision 0.1Date MAY 1996Page 20 of 37Table 4-1DCP ID AssignmentsIMPROVE Transmissometer NetworkDCP-ID CHAN LOCATION TIME * RATE **FA4

Strany 34 - Page 15 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 2 of 52.2 INSTRUMENT TECHNICIANThe instrument technician shall:• Perform all

Strany 35 - Page 16 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 21 of 37EXECUTE Execute the Handar DCP communications program "TERM" fromPROGRAM the PC comput

Strany 36 - Page 17 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 22 of 37• $ Forced Scan• # Forced Transmit (GOES Radio)Enter I to edit the station ID. The download pro

Strany 37 - Page 18 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 23 of 37Enter M to select DCP channel 01. The display message will be "PCHANNEL 01."Enter N to

Strany 38 - Page 19 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 24 of 37LABORATORY PERFORMANCE TESTINGHANDAR 540A/570A DCPDCP S/N: __________________________Date: _____

Strany 39 - Page 20 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 25 of 374.4.3 Run Mode Timing ChecksVERIFY Verify that the GOES primary channel select switches are set

Strany 40 - Page 21 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 26 of 37Set the power element of the wattmeter for the forward direction.Monitor the wattmeter reading a

Strany 41 - Page 22 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 27 of 37APPENDIX AHANDAR 540A DCP CONFIGURATION PROGRAM - 540ROT.DCP

Strany 42 - Page 23 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 28 of 37P ID FA43D0BEP STATION TIME 15:47:38P YEAR (XX) 94P DCP JULIAN DATE 055P GOES PRI XMT

Strany 43

Number 4250-2010Revision 0.1Date MAY 1996Page 29 of 37P CHANNEL NO. 03P03 SENSOR TYPE 10P03 CARD SLOT # 06

Strany 44 - Page 25 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 30 of 37P CHANNEL NO. 06P06 SENSOR TYPE 10P06 CARD SLOT # 06

Strany 45 - Page 26 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 3 of 5• Reference thermometer (°C)• Replacement components as required• Batte

Strany 46 - APPENDIX A

Number 4250-2010Revision 0.1Date MAY 1996Page 31 of 37P08 LOW DIFF LIMIT NO LIMITP CHANNEL NO. 09P09 SENSOR TYPE 10

Strany 47 - Page 28 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 32 of 37APPENDIX BHANDAR 570A DCP CONFIGURATION PROGRAM - 570TROT.DCP

Strany 48 - Page 29 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 33 of 37P ID FA43D0BEP STATION TIME 22:46:25P YEAR (XX) 94

Strany 49 - Page 30 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 34 of 37P02 MEAS INTERVAL 01:00:00P02 START OF MEAS 17:30:00P02 LEVEL 1 MEAS TYPE 001

Strany 50 - Page 31 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 35 of 37P05 SENSOR TYPE 10P05 SENSOR NAME TAG 10P05 CARD SLOT # 01

Strany 51 - APPENDIX B

Number 4250-2010Revision 0.1Date MAY 1996Page 36 of 37P07 ZERO SCALE 000P07 MEAS INTERVAL 01:00:00P07 START OF MEAS 17:30:00

Strany 52 - Page 33 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 37 of 37P CHANNEL NO. 10P10 SENSOR TYPE 12P10 SENSOR NAME TAG 12

Strany 53 - Page 34 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 4 of 5• Frequency counter• Standard electronics laboratory small tools• Solte

Strany 54 - Page 35 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 5 of 5• Routine laboratory servicing• DCP programming• Pre-field performance

Strany 55 - Page 36 of 37

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF CAMPBELL 21X DATALOGGERSTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Strany 56 - Page 37 of 37

Number 4250-2000Revision 1.0Date MAY 1996Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Mana