Number 4250-2000
Revision 1.0
Date MAY 1996
Page 3 of 7
Replacement components as required
Medium screwdrivers (flat-blade and Phillips-head)
Battery charger
Replacement sealed lead acid battery pack
Two (2) dry half-unit DESI PAK desiccant packets
4.0 METHODS
Campbell 21X dataloggers should be serviced according to the following schedule:
Prior to installation at a field monitoring site
On an annual schedule (for units not used at field sites)
Any time the operation or accuracy of the datalogger appears to be suspect
Calibration of the Campbell 21X datalogger is required any time calibration checks indicate
that the datalogger is not operating within factory specifications.
This section includes six (6) subsections:
4.1 Internal Memory Check
4.2 Analog Input Checks
4.3 Analog Output Checks
4.4 Pulse Counter Check
4.5 Panel Temperature Test
4.6 Internal Battery Servicing
4.7 Archiving Datalogger Service Records
Procedures for performing the internal memory check are documented on the Campbell
21X Datalogger Servicing Documentation Form (Figure 4-1) and are described in the following
sections.
RECORD Record the datalogger serial number and the current date. The
GENERAL initials of the technician performing the inspection should also be
INFORMATION recorded.
4.1 INTERNAL MEMORY CHECK
The Campbell 21X datalogger will perform an internal memory check on power-up. This
check indicates the status of each memory chip on the datalogger's CPU board. Procedures for
performing the internal memory check are documented on the Campbell 21X Datalogger
Servicing Documentation Form (Figure 4-1) and are as follows:
TURN Turn the datalogger
ON
. The datalogger display will read
DATALOGGER "HELLO."
ON