Section 7. Measurement Instructions
Half Bridge Parameters
Parameter
& Data Type
Enter
Dest
Variable or
Array
The Variable in which to store the results of the instruction. When Reps are used the
results are stored in an array with the variable name. An array must be dimensioned to
have elements for all the Reps.
Reps
Constant
The number of repetitions for the measurement or instruction.
Range
Constant
Alpha Code
Voltage Range
mV5000
± 5000 mV
mV2500
± 2500 mV
mV250
± 250 mV
mV25
± 25 mV
mV7_5
± 7.5 mV
mV2_5
± 2.5 mV
Autorange
mV2_5 mV5000
Selects range (Sect. 3.1)
mV250C
± 250 mV The mV250C, mV25C, mV7_5C, and
mV25C
± 25 mV mV2_5C ranges pull the channel into
mV7_5C
± 7.5 mV common mode range and check for open
mV2_5C
± 2.5 mV input
AutorangeC
mV2_5 mV250
Selects C range
SEChan
Constant
The single-ended channel number on which to make the first measurement. When Reps
are used, subsequent measurements will be automatically made on the following single-
ended channels. If the channel is entered as a negative number, all reps occur on that
channel.
ExChan
Enter the excitation channel number to excite the first measurement.
Constan
Alpha
Code
Code/
Channel
Result
VX1 1 Switched excitation channels, are switched to the excitation
voltage.
VX2 2 for the measurement and switched off between measurements.
VX3 3
MeasPEx
Constant
The number of sensors to excite with the same excitation channel before automatically
advancing to the next excitation channel. To excite all the sensors with the same
excitation channel, the number should equal the number of Reps.
ExmV
Constant
The excitation voltage in millivolts. Allowable range ± 2500 mV. RevEx may be used to
excite with both a positive and negative polarity to cancel offset voltages.
RevEx
Code Value Result (Reversing requires twice as much time to complete)
Constant False 0 Excite only with the excitation voltage entered
True
0
A second measurement is made with the voltage polarity
reversed to cancel offsets
RevDiff
Code Value Result (Reversing requires twice as much time to complete)
Constant False 0 Signal is measured with the high side referenced to the low
True
0
A second measurement is made after reversing the inputs to cancel
offsets
7-7