SECTION 11. OUTPUT PROCESSING INSTRUCTIONS
11-6
PAR. DATA
NO. TYPE DESCRIPTION
01: 2 Storage area option
01 = Final Storage (00 and
02 also default to
Final Storage)
03 = Input Storage Area
02: 4 Starting input location
destination if option 03
Output Array ID if options 0-2
(1-511 are valid IDs)
*** 82 STANDARD DEVIATION IN TIME ***
FUNCTION
Calculate the standard deviation (STD DEV) of
a given input location. The standard deviation
is calculated using the formula:
S = ((ΣX
i
2
- (ΣX
i
)
2
/N)/N)
1/2
where X
i
is the ith measurement and N is the
number of samples.
PAR. DATA
NO. TYPE DESCRIPTION
01: 2 Repetitions
02: 4 Starting input location number
Outputs generated: 1 per repetition