Appendix B. Entering SDI-12 Commands/Changing Settings
TABLE B-1. SDI-12 Commands
Function SDI-12 Command
Address Query
?!
Send Identification
aI!
Acknowledge Active
a!
Change Address
aAb!
Where a is the current address and b is the
new address
Start Verification
aV!
Start Measurement
aM!
Start Measurement and
Request CRC
aMC!
Send Data
aD0!...aD9!
Additional Measurements
aM1!...aM9!
Additional Measurement and
Request CRC
aMC1!...aMC9!
Start Concurrent Measurement
aC!
Start Concurrent Measurement
and Request CRC
aCC!
Additional Concurrent
Measurements
aC1!...aC9!
Additional Concurrent
Measurements and Request
CRC
aCC1!...aCC9!
Start False Echo Learn
aXSFEL+n!
n=distance necessary to clear the obstruction
Set Power n Operation Mode
(ON, OFF, AUTO)
aXSPOM+n!
n=2 (auto), 1 (on), or 0 (off)
Get Power Operation Mode
(ON, OFF, AUTO)
aXGPOM!
2=auto; 1=on; 0=off
Get Water Conditions
aXGWC!
n=0 (auto), 1 (smooth) 2 (medium), or 3
(rough)
Set Water Conditions
Where n is the new Water
Condition
aXSWC+n!
n=0 (auto), 1 (smooth) 2 (medium), or 3
(rough)
Set Water Stage
Where n is the new Water
Stage
aXSS+n!
n=floating point number that is the initial
distance between the sensor and the water
surface.
Get Unit
aXGU!
0=meters; 1=feet
Set Unit
Where n is the new unit of
measurement
aXSU+n!
n= 0 (meters) or 1 (feet)
Reset Sensor
Resets the sensor to its factory
settings
aXRS!
0=reset unsuccessful; 1=reset successful