Shrnutí obsahu

Strany 1

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer2/12Copyright © 2010-2012Campbell Scientific, Inc.

Strany 2

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer o The SDM, USB, and RS-485 output options include IRGASON diagnostic data.

Strany 3

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer • Standard outputs: o ux, uy, and uz orthogonal wind components o sonic t

Strany 4

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Sonic accuracy5 Offset error ux, uy: < ±8 cm·s-1 uz: < 4 cm·s-1 Gai

Strany 5

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 5.2 Output Signals Features • EC100 electronics outputs data using: o CS

Strany 6

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer FIGURE 5-1(a). IRGASON dimensions, side view FIGURE 5-1(b). IRGASON dime

Strany 7

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 5.4 Power Requirements During Power-up and Operation Power requirement Sour

Strany 8

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Handle the IRGASON carefully. The optical source may be damaged by rough h

Strany 9

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer IRGASON CM250 Leveling Mount (p/n 26559) IRGASONMounting Bracket(p/n 27496)

Strany 10

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer FIGURE 6-2. EC100-enclosure mounting bracket mounted on a vert

Strany 11

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 8. Mount the IRGASON temperature probe so that it measures temperature at t

Strany 13

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer tighten the thumbscrews (see Figure 6-5). The gas analyzer cable is approxi

Strany 14

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer FIGURE 6-6. Base of EC100 enclosure Unique calibration information is stor

Strany 15

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 3. Connect the IRGASON temperature probe. Unscrew the temperature connecto

Strany 16

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer o Analog output: use cable CABLE4CBL-L (p/n 21972) or CABLE2TP-L (p/n 26986

Strany 17

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 7.1 Factory Defaults Table 7-1 shows the default value for each setting des

Strany 18

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer any integer value between 0 and 14. The value 15 is reserved as an SDM-group

Strany 19

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer troubleshooting only. In normal operation, Temperature Sensor is set to Aut

Strany 20

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 7.14 ECMon Settings for the IRGASON are easily verified or changed using th

Strany 21

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer FIGURE 7-1. Main window of ECMon software 20

Strany 22

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer FIGURE 7-2. Setup window in ECMon 7.15 Device Configuration Utility DevCo

Strany 23

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Strany 24

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 8. EC100 Outputs The EC100 is the IRGASON electronics module that measures

Strany 25

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer To use USB or RS-485 output, connect a USB or RS-485 cable from the EC100 to

Strany 26

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer The final data element in each row or output array is the signature. The si

Strany 27

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer TABLE 8-2. Multipliers and Offsets for Analog Outputs Density (mg m-3) Volt

Strany 28

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Remove the wicks during the winter. The wicks will accumulate snow or freez

Strany 29

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Wicks are available as listed in Table 9-1. TABLE 9-1. Rain Wick Replacement

Strany 30

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 9.3 Gas-Analyzer Window Cleaning The windows of the analyzer should be clea

Strany 31

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer rubber seals on the ends of the shroud are clean and in good condition. Pos

Strany 32

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 8. Connect a gas tank of known CO2 concentration to a pressure regulator, t

Strany 33

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 10. Remove the H2O span gas, and connect a zero air source (no CO2 or H2O)

Strany 34

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strany 35

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 16. Enter the known dew point (in C°) in the box labeled Span Dew Point and

Strany 36

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Scrubber bottle cover Scrubber bottle cover FIGURE 9-5. Replacing the desi

Strany 37

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 10. Datalogger Programming The datalogger is programmed in the CRBasic lang

Strany 38

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer TABLE 10-1. Output Modes for EC100 Instruction OutputMode Data Field Desc

Strany 39

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer TABLE 10-3. Bits in the Gas Diagnostic Flag bit hex value decimal Na

Strany 40

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 10.2 EC100Configure() Instruction EC100Configure() provides a third option,

Strany 41

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer TABLE 10-4. ConfigCmd Values for Setting and Retrieving Settings ConfigCmd

Strany 42

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 10.2.1 ConfigCmd 11 Zero-and-span Control To perform zeroing of CO2 and H2O

Strany 43

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer output indicates moisture is present, the datalogger can send the EC100Confi

Strany 44

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 11. Theory of Operation 11.1 IRGASON Sonic Anemometer The CSAT3 measures w

Strany 45

IRGASON Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links

Strany 46

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 11.1.2 Temperature The sonically determined speed of sound can be found fro

Strany 47

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer The sonic virtual temperature, in degrees Celsius, is given by Eq. (9), wher

Strany 48

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 44 11.3 Theory References Kaimal, J. C. and Businger, J. A.: 1963, “A Cont

Strany 49

Appendix A. Filter Bandwidth and Time Delay The EC100 measures CO2, H2O, 3-D wind components, and sonic temperature from the IRGASON at 100 Hz and th

Strany 50

Appendix A. Filter Bandwidth and Time Delay The EC100 measures gas and wind data at 100 Hz, and 100 Hz data are down-sampled to the datalogger scan r

Strany 51

Appendix A. Filter Bandwidth and Time Delay 110Hertz700.00010.0010.010.1110No UnitsEC100 10-Hz Bandwidth Filter10-Hz Bandwidth from a 50-msec Moving

Strany 52

Appendix A. Filter Bandwidth and Time Delay A-4

Strany 53

Appendix B. Equations Table B-1 lists variables and constants used in the following equations. TABLE B-1. Table of Equation Variables and Constants

Strany 54

Appendix B. Equations ()⎟⎟⎠⎞⎜⎜⎝⎛+−⎟⎟⎠⎞⎜⎜⎝⎛+=vvddXXTRPM1000115.273ρ (B-5) Dew Point from Molar Mixing Ratio tmpdtmpddTTT__502.17 97.240−= (B-62) ()⎟⎟

Strany 55

Appendix C. Material Safety Data Sheets (MSDS) MSDS are available for chemicals used in IRGASON filters. The MSDS samples below are made available f

Strany 56

IRGASON Table of Contents 8. EC100 Outputs ...22 8.1 SDM Output...

Strany 58

IRGASON Table of Contents 6-4. Mounting the IRGASON Temperature Probe...11 6-5. EC100 electronics front panel wi

Strany 59

IRGASON Table of Contents iv

Strany 60

IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 1. Introduction The IRGASON is an in-situ, open-path, mid-infrared absorpti