Campbell CDM-VW300 Series Manuál Strana 115

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 122
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 114
Appendix G. CRBasic Program Library
G.1.14 100 Hz Measurement Example — One CDM-VW300, Two
Channels
'===100Hz-1Device2Ch_3-25-13.CR3===
'CR3000 datalogger
'CDM-VW300 vibrating-wire analyzer
'Program to read 100-Hz dynamic data from one CDM-VW300 analyzer measuring two channels
'IMPORTANT -- Ensure that the CPI address coded on the following line matches the address
'reported for the attached analyzer in the DevConfig or DVWTool software.
Const CPI_ADDR = 1 '<<<<<<<<<<<<<<SET CPI ADDRESS HERE
Public Freq(2) 'dynamic frequencies
Public Diag(2) As Long 'diagnostic code
Public StaticFreq(2) 'Static (1Hz output) frequencies
Public Therm(2) 'Thermistor readings
'Standard Deviation of the dynamic readings that occurred during the latest one-second interval.
Public DynStdDev(2)
'The following arrays are used to configure the CDM-VW300 series device. Refer to the
'CDM_VW300Config instruction used below
'
' CH1 CH2
' --- ---
'Set to true (Enabled=1, Disabled=0) only those channels which have sensors connected
Dim Enable(2) As Long = { 1, 1}
'Specify the target/desired resonant amplitude at which the sensor will be maintained
'via excitation, given in Volts. This should be in the range 0.010 to 0.001
Dim Max_AMP(2) = { 0.002, 0.002}
'Low Frequency Boundary (sensor frequency should never fall below
'this value regardless of environmental changes)
Dim F_Low(2) = { 300, 300}
'High Frequency Boundary (sensor frequency should never exceed
'this value regardless of environmental changes)
Dim F_High(2) = { 6000, 6000}
'Output Format - Hz vs. Hz^2 :: Value of 0 – measured frequency is given in units of Hz,
'Value of 1 – measured frequency is squared and given in units of Hz^2
Dim OutForm(2) As Long = { 0, 0}
'Multiplier (factor) to be applied to sensor output frequency
Dim Mult(2) = { 1.0, 1.0}
'Offset (shift) to be applied to sensor output frequency
Dim Off(2) = { 0.0, 0.0}
'Steinhart-Hart coefficients [A,B,C] for converting thermistor ohms to
'temperature in Celsius. Specifying zeroes for A,B,C results in a reading in Ohms.
Dim SteinA(2) = { 0.0, 0.0}
Dim SteinB(2) = { 0.0, 0.0}
Dim SteinC(2) = { 0.0, 0.0}
'Rainflow configuration (not used in this program, but required as configuration arguments)
Dim RFMB(2) As Long = { 20, 20}
Dim RFAB(2) As Long = { 20, 20}
Dim RFLL(2) = { 400.0, 400.0}
Dim RFHL(2) = {4000.0,4000.0}
Dim RFHY(2) = { 0.005, 0.005}
Dim RFOF(2) As Long = { 100, 100}
'Configure the CDM-VW300 series device. Use the variable arrays declared above.
CDM_VW300Config(0,CPI_ADDR,0,Enable(),Max_AMP(),F_Low(),F_High(), _
OutForm(),Mult(),Off(), SteinA(),SteinB(),SteinC(), _
RFMB(),RFAB(),RFLL(),RFHL(),RFHY(),RFOF())
DataTable (static,true,-1)
'Static Frequency reading (1Hz output)
Sample (2,StaticFreq(),IEEE4)
'Thermistor reading : Ohms or DegC
Sample (2,Therm(),IEEE4)
'Standard Deviation of dynamic readings
'taken during the most recent second
Sample (2,DynStdDev(),IEEE4)
EndTable
DataTable (dynamic,true,-1)
'Dynamic Frequency (100Hz output)
Sample (2,Freq(),IEEE4)
'Diagnostic code for the current dynamic reading
G-23
Zobrazit stránku 114

Komentáře k této Příručce

Žádné komentáře