Campbell CDM-VW300 Series Manuál Strana 49

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 122
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 48
CDM-VW300 Series Dynamic Vibrating-Wire Analyzers
vibrating-wire sensors operate equally well with either polarity position of the
two signal wires. Two coil connections (C) are provided on each channel on
the CDM-VW300 to which these leads are connected.
FIGURE 7-6. Three-wire vibrating-wire sensor connection
The third lead is the shield wire, or sensor ground, and should be connected to
the ground connection corresponding to the measurement channel.
Coil
Coil
Shield / Sensor Ground
Vibrating-Wire Sensor
Thermistor
Thermistor
FIGURE 7-7. Five-wire, vibrating-wire sensors leads
For sensors with five lead wires, three wires operate as described previously
for a three-wire configuration. The additional two leads are thermistor leads,
and, as illustrated in FIGURE 7-7 and FIGURE 7-8, can be wired to the
thermistor (T) channels on the analyzer.
39
Zobrazit stránku 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 121 122

Komentáře k této Příručce

Žádné komentáře