Strany 1 - INSTRUCTION MANUAL

INSTRUCTION MANUALTDR100Revision: 9/05Copyright (c) 2000-2005Campbell Scientific, Inc.

Strany 2 - Warranty and Assistance

TDR100 4 1 is the default setting and can be changed in PCTDR menu Settings/Communications. The baud rate is factory set to 57600. 3. Connect 12 vol

Strany 3 - TDR100 Table of Contents

TDR100 5 FIGURE 3-1. Waveform of a CS610 in water. Changing the Waveform Start value to 5.7 m and the Waveform Length to 5 m gives the waveform disp

Strany 4

TDR100 6 FIGURE 3-2. Waveform of CS610 in water after changing Start and Length parameters to display relevant portion of reflected signal. 3.2 PCT

Strany 5

TDR100 7 When the TDR100 is connected to a CR23X or CR1000 datalogger using control ports 1-3 for SDM control and SDMX50 multiplexers are also connect

Strany 6

TDR100 8 4.3 Menu Selections 4.3.1 File Menu Save Configuration/Load Configuration - save and reload configuration of user-selectable parameters. S

Strany 7 - 1. Introduction

TDR100 9 4.4.2 Waveform Average - sets the number of measurements averaged at a given distance from the TDR100. A value of 4 is recommended. Higher

Strany 8 - 2. System Specifications

TDR100 10 FIGURE 4-1. PCTDR waveform for CS610 in water. The algorithm will use the length of the waveform set by the Waveform Length. After findin

Strany 9

TDR100 11 ()LLwv=⋅++−θmax..017601142 with L the actual probe rod length, and, θv-max the maximum expected volumetric water content. Two m is added for

Strany 10

TDR100 12 LoggerNet (version 3.0 or higher) are used to create and send the CRBasic Program to the CR1000 datalogger. 5.3 TDR100 The TDR100 contains

Strany 11

TDR100 13 5.4.1 Grounding The TDR system should be installed with a single ground point. A good earth ground should be established close to the data

Strany 12 - 4. PCTDR Software

Warranty and Assistance The TDR100 is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and

Strany 13 - 4.2 PCTDR Help

TDR100 14 TABLE 5-1. SDM Addressing Base 10 Number Instruction 119 Address (Base4) TDR100 Thumbwheel Switch Level 1 SDMX50 Mux (TDR100+1) MSD LSD L

Strany 14 - 4.3 Menu Selections

TDR100 15 5.6.1 Mounting Equipment in ENCTDR100 The ENCTDR100 is a 16” x 18” weathertight enclosure that is modified for use with a Campbell Scientif

Strany 15 - 4.4.2 Waveform

TDR100 16 from voltage surge damage. Figure 5-3 describes typical signal attenuation from the common to one output channel for frequency range import

Strany 16

TDR100 17 FIGURE 5-4. Location of Address Jumpers on SDMX50 5.8 Soil Probes The TDR probes are the sensors of the TDR measurement system and are in

Strany 17 - 5.2 Datalogger

TDR100 18 5.8.1 Determining Probe Constant, Kp, using PCTDR Section 6.2.6 presents the principles for TDR measurements of soil electrical conductivit

Strany 18 - 5.3 TDR100

TDR100 19 6.2 Discussion of TDR Instruction Parameters (Instruction 119) 6.2.1 Parameter 1: SDM Address The SDM address of the TDR100 is set by sele

Strany 19 - 5.5 SDM Communication

TDR100 20 Section 7 for discussion of the electrical conductivity measurement method and Section 6.2.13.1 for discussion of Kp.. 6.2.3 Parameter 3: M

Strany 20 - 5.6 ENCTDR100

TDR100 21 6.2.5 Parameter 5: Relative Propagation Velocity Relative propagation velocity is the ratio of actual propagation velocity for a transmissi

Strany 21 - 5.7 SDMX50

TDR100 22 6.2.10 Parameter 10: Probe Offset (meters) Most TDR probes have a block of epoxy or other material which holds the rods rigidly spaced and

Strany 22 - SDMX50 attenuation (dBm)

TDR100 23 6.2.12 Parameter 12: Multiplier Multiplication factor applied to the value stored in the input location specified in parameter 11. Set to

Strany 23 - 5.8 Soil Probes

i TDR100 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab

Strany 24 - , using PCTDR

TDR100 24 Entering these constants as the multiplier when using Instruction 119 for electrical conductivity measurement will give units of Siemens/met

Strany 25

TDR100 25 Mux/ProbeSelect: The Mux/Probe Select parameter is used to define the setup of any multiplexers and attached probes in the system. The addre

Strany 26

TDR100 26 parameter only has an affect when Option 0, La/L, is used for the measurement. Mult, Offset: The Mult and Offset parameters are each a const

Strany 27 - 6.2.6 Parameter 6: Points

TDR100 27 caused by the probe beginning and end. This information is then analyzed to determine soil water content. While the velocity of the applied

Strany 28 - 8 multiplier

TDR100 28 16 meter cable26 meter cable45 meter cable66 meter cable FIGURE 8-1. Waveforms collected in a sandy loam using CS610 probe with RG8 connect

Strany 29

TDR100 29 water content = 9.5%water content = 25% FIGURE 8-2. Waveforms collected in a sandy loam using CS610 probe with RG8 connecting cable. Volum

Strany 30 - Remarks

TDR100 30 9. Programming Examples 9.1 CR10X/CR23X Program Examples Example 1, Measure and Record; Analog Measurements and Volumetric Water Content A

Strany 31

TDR100 31 ;Turn off the switched 12V to power off the TDR100: 5: Do (P86) 1: 55 Set Port 5 Low ;Set the Output Flag to output data every 5 m

Strany 32 - 7. TDR Principles

TDR100 32 Example 2, Measure and Record; Analog Measurements, Volumetric Water Content, Electrical Conductivity, and Capture a Waveform A CS605 or CS6

Strany 33

TDR100 33 ;Measure the CS610 probe connected to Channel 3 of the level 1 SDMX50 and return La/L. 5: TDR100 Measurement (P119) 1: 0 SDM Address 2

Strany 34 - Measurement

TDR100 Table of Contents ii 5.8 Soil Probes ... 17 5.8.1 D

Strany 35

TDR100 34 ;Measure Electrical Conductivity on the same CS610 TDR probe connected to Channel 3. 9: TDR100 Measurement (P119) 1: 0 SDM Address 2:

Strany 36 - 9. Programming Examples

TDR100 35 ;Output the Battery_Voltage, DataLogger Temperature, WaterContent, ;and Electrical Conductivity. 14: Sample (P70) 1: 4 Reps 2: 1

Strany 37

TDR100 36 Example 3, Measure and Record; Volumetric Water Content and Analog Measurements Eight CS605 or CS610 probes are connected to Channel #1 thro

Strany 38

TDR100 37 4: Beginning of Loop (P87) 1: 0 Delay 2: 8 Loop Count ;Square La/L to convert to dielectric constant: ;Note: The user must manually

Strany 39 - (meters)

TDR100 38 ;Output the 8 water content measurements and battery voltage: 12: Sample (P70) 1: 9 Reps 2: 11 Loc [ WC_1 ] *Table 2 Progra

Strany 40

TDR100 39 Example 4, Measure and Record; Analog Measurements and Volumetric Water Content In this example analog measurements are made every 5 minutes

Strany 41

TDR100 40 5: TDR100 Measurement (P119) 1: 00 SDM Address 2: 0 La/L 3: 3006 MMMP Mux & Probe Selection 4: 4 Waveform Averaging 5:

Strany 42 - Measurements

TDR100 41 ;Switch the level 1 SDMX50 to Channel 2, the level 2 SDMX50s to Channel 8 and ;measure the 8 probes (channels 1 through 8) on the Level 3 S

Strany 43

TDR100 42 ;Output the 29 TDR water content measurements: 15: Sample (P70) 1: 29 Reps 2: 11 Loc [ WC_3001 ] *Table 2 Program 02: 0.0000

Strany 44

TDR100 43 Example 5, Trouble Shooting Program for SDMX50 Multiplexer, Datalogger Control Ports, and Analog Measurements The following program can be u

Strany 45 - Water Content

TDR100 Table of Contents iii 8-1 Waveforms collected in a sandy loam using CS610 probe with RG8 connecting cable. Volumetric water content is 24% an

Strany 46

TDR100 44 9.2 CR1000 Program Example Example 1, Measure and Record; Analog Measurements, Volumetric Water Content, and Capture TDR Probe Waveforms Eq

Strany 47

TDR100 45 Public Flag(2) public WavePT(260) public MuxChan Dim I 'Declare Constants ------------------------------------------------ 'Topp

Strany 48

TDR100 46 If TimeIntoInterval(0,24,Hr) then flag(2)=high ' If Flag(1)=High Then '************************************* SW12 (1

Strany 49

TDR100 47 TDR100 (WavePT(),0,1,6001,4,1.0,251,9.5,5.0,0.3,0.085,1,0) CallTable TDR_Wave() ' MuxChan=7001 TDR100 (WavePT(),0,1,7001

Strany 50 - 9.2 CR1000 Program Example

TDR100 48 This is a blank page.

Strany 51

This is a blank page.

Strany 52

Campbell Scientific CompaniesCampbell Scientific, Inc. (CSI)815 West 1800 NorthLogan, Utah 84321UNITED [email protected]

Strany 53 - 10. References

TDR100 Table of Contents iv This is a blank page.

Strany 54 - This is a blank page

1 TDR100 1. Introduction This document presents operating instructions for the TDR100 and associated equipment and discusses time domain reflectometr

Strany 55

TDR100 2 2. System Specifications 2.1 General See the CR10X, CR23X, or CR1000 datalogger manuals for datalogger specifications. 2.2 Power Consumpti

Strany 56 - Campbell Scientific Companies

TDR100 3 waveform averaging 1 to 128 electrostatic discharge protection internal clamping power supply unregulated 12 volt (9.6 V to 16 V), 300 mil