TDR100
14
TABLE 5-1. SDM Addressing
Base 10
Number
Instruction 119
Address
(Base4)
TDR100
Thumbwheel
Switch
Level 1
SDMX50 Mux
(TDR100+1)
MSD LSD
Level 2
SDMX50 Muxs
(TDR100+2)
MSD LSD
Level 3
SDMX50 Muxs
(TDR100+3)
MSD LSD
0 00 0 0 1 0 2 0 3
1 01 1 0 2 0 3 1 0
2 02 2 0 3 1 0 1 1
3 03 3 1 0 1 1 1 2
4 10 4 1 1 1 2 1 3
5 11 5 1 2 1 3 2 0
6 12 6 1 3 2 0 2 1
7 13 7 2 0 2 1 2 2
8 20 8 2 1 2 2 2 3
9 21 9 2 2 2 3 3 0
10 22 A 2 3 3 0 3 1
11 23 B 3 0 3 1 3 2
12 30 C 3 1 3 2 3 3
13 31 D 3 2 3 3 not available
14 32 E 3 3 not available not available
15 33 F not available not available not available
5.5.2 SDM Cable and Cable Length Considerations
5-conductor cable with shield and drain is used for interconnection of SDM
devices. The 5 conductors are used for 12 volt power, ground and the 3 SDM
lines. A cable assembly (pn13776) is provided with the TDR100 and the
ENCTDR100. This assembly is for SDM connection between the TDR100
and a datalogger and between the TDR100 and a SDMX50 multiplexer.
Additional cable for SDM connection to other TDR system components is
available from Campbell Scientific as part SDMCBL-L with -L the cable
length in feet. The maximum recommended total length of all SDM cables
should not exceed 500 feet. Lengths greater than 500 feet can give unreliable
communication between SDM devices.
SDM communications use datalogger control ports 1, 2 and 3. No other
devices should be connected to these control ports.
The insulation for the individual wires of the SDM cable affects the frequency
response and reliability. PVC insulation has more attenuation than
polypropylene or polyurethane and should not be used for SDM
communication except when total SDM cable length is less than 250 feet.
5.6 ENCTDR100
Many TDR system applications require installation of equipment at field sites.
The ENCTDR100 is a weather-proof enclosure with a mounting plate for a
CR10X, CR23X, or CR1000 datalogger, power supply, TDR100, SDMX50SP,
cable strain relief bracket and associated cabling. The ENCTDR100 can be
mounted on a CM6 or CM10 tripod for free-standing installation.