TDR100
33
;Measure the CS610 probe connected to Channel 3 of the level 1 SDMX50 and return La/L.
5: TDR100 Measurement (P119)
1: 0 SDM Address
2: 0 La/L
3: 3001 MMMP Mux & Probe Selection
4: 4 Waveform Averaging
5: 1 Vp
6: 250 Points
7: 9.5 Cable Length (meters)
8: 5 Window Length (meters)
9: .3 Probe Length (meters)
10: .085 Probe Offset (meters)
11: 3 Loc [ LaL_CH3 ]
12: 1.0 Mult
13: 0.0 Offset
;Square La/L to convert to dielectric constant:
6: Z=X*Y (P36)
1: 3 X Loc [ LaL_CH3 ]
2: 3 Y Loc [ LaL_CH3 ]
3: 4 Z Loc [ WC_CH3 ]
;Multiply the dielectric constant by 0.1 to prepare for the 3rd order polynomial
;with the appropriately scaled coefficients for Topp's calibration:
7: Z=X*F (P37)
1: 4 X Loc [ WC_CH3 ]
2: 0.1 F
3: 4 Z Loc [ WC_CH3 ]
;Apply Topp's 3rd order polynomial:
8: Polynomial (P55)
1: 1 Reps
2: 4 X Loc [ WC_CH3 ]
3: 4 F(X) Loc [ WC_CH3 ]
4: -0.053 C0
5: .292 C1
6: -0.055 C2
7: 0.0043 C3
8: 0.0 C4
9: 0.0 C5