Using Campbell AVW2xx Vibrating Wire Interfaces – MultiLogger Application Note #21 – Page 6 of 13
Wired Configuration – AVW200 with Direct Connect Gages
The following configuration depicts an AVW200 connected directly to the CR800 or CR1000, with gages
connected directly (no multiplexers) to the AVW200. Up to 2 gages may be directly connected to the
AVW200. The connection scheme utilizes SDI-12, this allows a single control port connection to the MCU.
Other configurations are possible including use of the RS-232 port. See the AVW200 Operators Manual
for more information on these configurations.
To configure MultiLogger use the following Direct Connect Gage Type | Make | Model selections:
Units
Vibrating Wire Generic AVW200 CH1 12V Digits Channel 1 with 12V excitation Digits
AVW200 CH1 12V Hz Channel 1 with 12V excitation Hz
AVW200 CH1 5V Digits Channel 1 with 5V excitation Digits
AVW200 CH1 5V Hz Channel 1 with 5V excitation Hz
AVW200 CH2 12V Digits Channel 2 with 12V excitation Digits
AVW200 CH2 12V Hz Channel 2 with 12V excitation Hz
AVW200 CH2 5V Digits Channel 2 with 5V excitation Digits
AVW200 CH2 5V Hz Channel 2 with 5V excitation Hz
Notes:
• All of these Models use a frequency range of 400-4000Hz which should be adequate for most
installations. Contact Canary Systems for direction to narrow or widen the frequency range.
• All of these Models use SDI-12 address 0. Contact Canary Systems for direction when using a
different SDI-12 address.
To read the Upper Channel (Temperature Device):
Units
AVW200_YSI44005-°C YSI type 44005 on Channel 1 or 2 °C
AVW200_YSI44005-°F YSI type 44005 on Channel 1 or 2 °F
1V+
1V-
1T+
1T-
GND
2V+
2V-
2T+
2T-
GND
G
C1 (SDI12)
C2 (CLK)
C3 (RES)
G
G
12V
C7 (CR1000)
G
12V
vary by manufacturer.
RED
BARE
BLACK
WHITE
GREEN
VW#1
RED
BARE
BLACK
WHITE
GREEN
VW#2