
AM16/32B Relay Analog Multiplexer
EXAMPLE “2x32” MODE PROGRAMS - GENERALIZED PROGRAM LOOPS FOR THE
CR23X, 21X, CR10(X), AND CR7.
21X SAMPLE PROGRAM
* 1 Table 1
Programs
01: 60 Sec.
Execution
Interval
;ACTIVATE MULTIPLEXER
1: Set Port (P20)
1: 1 Set high
2: 1 Port
Number
;BEGIN MEASUREMENT
;LOOP
2: Beginning of Loop (P87)
1: 0 Delay
2: 32 Loop Count
CLOCK PULSE/DELAY
3: Excitation with delay (P22)
1: 1 EX Chan
2: 1 Delay w/EX
(units=
.01 sec)
3: 1 Delay after
EX (units=
.01 sec)
4: 5000 mV
Excitation
4: User Specified Measurement
Instruction
;END MEASUREMENT
;LOOP
5: End (P95)
;DEACTIVATE
;MULTIPLEXER
6: Set Port (P20)
1: 0 Set low
2: 1 Port
Number
CR7 SAMPLE PROGRAM
* 1 Table 1
Programs
01: 60 Sec.
Execution
Interval
;ACTIVATE MULTIPLEXER
1: Set Port (P20)
1: 1 Set high
2: 1 EX Card
3: 1 Port No.
;BEGIN MEASUREMENT
;LOOP
2: Beginning of Loop (P87)
1: 0 Delay
2: 32 Loop Count
;CLOCK PULSE/DELAY
3: Excitation with delay (P22)
1: 1 EX Chan
2: 2 EX Chan
3: 1 Delay w/EX
(units=
.01 sec)
4: 1 Delay after
EX (units =
.01 sec)
5: 5000 mV
Excitation
4: User Specified Measurement
Instruction
;END MEASUREMENT
;LOOP
5: End (P95)
;DEACTIVATE
;MULTIPLEXER
6: Set PortP20
1: 0 Set low
2: 1 EX Card
3: 1 Port No.
CR10(X), CR23X SAMPLE
PROGRAM
* 1 Table 1
Programs
01: 60 Sec.
Execution
Interval
;ACTIVATE MULTIPLEXER
1: Do (P86)
1: 41 Set high
Port 1
;BEGIN MEASUREMENT
;LOOP
2: Beginning of Loop (P87)
1: 0 Delay
2: 32 Loop Count
;CLOCK PULSE
3: Do (P86)
1: 72 Pulse Port 2
;DELAY
4: Excitation with Delay (P22)
1: 1 EX Chan
2: 0 Delay w/EX
(units=.01 sec)
3: 1 Delay after EX
(units=.01 sec)
0: 0 mV Excitation
5: User Specified Measurement
Instruction
;END MEASUREMENT
;LOOP
6: End (P95)
;DEACTIVATE
;MULTIPLEXER
7: Do (P86)
1: 51 Set low
Port 1
FIGURE 9. Example “2x32” Mode Program Loops for CR23X, CR10(X), 21X and CR7 Loggers
22
Komentáře k této Příručce