APPENDIX F. DYNAGAGE SAP-FLOW (P67)
F-3
Appendix A
Instruction P67 Processing
Loc = input location assuming input locations 1,
2, 3, and 4 are used.
Par = Instruction P67 parameter.
P
in
= (Loc 4)*(Loc 4)/(Par 3)
Q
v
= (((Loc 3)-(Loc 2))/(4.0*(Par 6)))*(Par
4)*(Par 5)
Q
r
= (Loc 1)*(Par 2)
Q
f
= P
in
- Q
v
- Q
r
K
shapp
= (P
in
- Q
v
)/(Loc 1)
dT = ((Loc 2+Loc 3)/2.0)*25.0
Sapflow = Q
f
/(dT*4.186)
If Par 7 µ 0.0 then go to XXXXXX
If Q
f
< (0.2*P
in
) and If dT < Par 7, then Sapflow
= 0.0
If Q
f
< (0.2*P
in
) and If Q
f
< 0.0, then Sapflow =
- 0.00001
XXXXXX
If Par 8 µ 0.0 then go to YYYYYY
F
max
= (Par 8)*(Par 4)
If Sapflow > F
max
, the Sapflow = F
max
YYYYYY
Sapflow = (Sapflow*(Par 11))+(Par 12)
Transfer to Input Locations beginning with
Parameter 10, either short output or long output,
depending on Parameter 9.